Compact single-shot CEP-meter

Compact single-shot CEP-meter

Madeira (UltraFast Innovations) is a single-shot phasemeter that facilitates single-shot measurements of the carrier-envelope-phase (CEP) of few-cycle femtosecond laser pulses. The ability to characterize the CEP of few-cycle pulses is the backbone of attosecond metrology and key to achieving sub-cycle temporal resolution in a broad range of ultrafast laser applications.

This new device does not require high vacuum or multichannel plates, which guarantees simple and user-friendly operation. MADEIRA supports center wavelengths ranging from the visible to the near-infrared domain, and operates at a repetition rate of up to 10 kHz.

MADEIRA relies on the detection of the light-induced, CEP-dependent current. The current is generated via strong-field ionization of gas-phase ethanol in the focus of the laser pulse and is probed by a pair of electrodes placed on either side of the focus. The current flowing between the two electrodes is amplified and converted into a voltage output signal. A second output signal is obtained from a similar measurement of a phase-shifted copy of the pulse in a second focus. The CEP is retrieved from the two output signals, which can be acquired continuously without time limitation.

Key Product Features

  • Wavelength range: 500-1000 nm
  • Pulse duration: <=4.5fs @750 nm central wavelength
  • Input polarization: linear, p-pol
  • Input pulse energy: 10-40 uJ, best performance at 30 uJ
  • Repetition rate: 10 kHz
  • Footprint: 35 x 30 cm2
  • Phase resolution: down to 200 mrad

More information :
https://www.ultrafast-innovations.com/devices/MADEIRA.html
https://www.optonlaser.com/fr/produit/mesure-de-cep-d-impulsions-uniques